UV Preconditioning Chamber
Meet the UV preconditioning test requirements specified in the 4.10 clause of the IEC 61215-2016 standard and the standard MST54 of the IEC 61730-2016 standard. Specially designed UV metal halide lamps are used to test of photovoltaic products and materials.
l Temperature of lamp could be by cooled down in closed chamber which is cooled by water cooling chiller .
l Module temperature could be controlled by water cooling chiller with cooling air blowing the module.
l UV metal halide lamps are used with high value performance, low test cost.
l The irradiance of each UV lamp can be controlled separately to adjust the irradiance of each lamp.
l During the test, the UV irradiance data can be tracked. After the irradiation reaches the requirement, the test can be stopped automatically.
l With the power off memorizing function, which could accumulate on the data before power off.
l Real-time curves of UVA & UVB radiation, temperature of module could be showed on the display and history curves could be accessed.
l Data of UVA & UVB radiation, temperature of module could be could be memorized in the hard disc of the IPC as EXCEL format.
Item  | Parameters  | 
Uniformity of irradiation surface  | ≤15%  | 
Spectral range  | 280nm~400nm  | 
UV irradiance(280nm-400nm)  | 150~250W/m2  | 
UVB ratio  | 3%~10%  | 
UVA Radiometer Range  | 0~300W/m2  | 
UVA Radiometer accuracy  | ±3%  | 
UVB Radiometer Range  | 0~50W/m2  | 
UVB Radiometer accuracy  | ±3%  | 
Temperature of the module  | 60±5℃  | 
Temperature accuracy  | ±2℃  | 
Temperature repeatability  | ±0.5℃  | 
Lamp area air temperature  | 25±5℃  | 
Temperature quantity  | 2  | 
Light Area  | 2200 *1200 mm  | 